Chapter 6 Diagnostics/Maintenance
220
Figure 104 Barrier Exerciser Screen
Selectable Line Items
Cycles sets how many open close cycles should be done. Default is 1. Maximum is
1,000,000. It can be edited via the Ctrl+E = Edit function.
Barrier Opened Time sets how long barrier is kept open before closing it again.
Barrier Closed Time set how long barrier is kept closed before opening it again.
Total Hot Cycles Done and Total Cold Cycles Done counts: These counts
accumulate over multiple F1 - Start cycles.
Hot Side (No or Yes) determines whether Hot Side Barrier is exercised.
Cold Side (No or Yes) determines whether Cold Side Barrier is exercised.
Non-Selectable Line Items
Status Information for several parameters as described by their labels.
Function Keys
F1 = Start: Starts the specified number of cycles.
F3 = Stop: Stops before the specified number of cycles is complete.
Ctrl + F4 = Clear Total Cycles Done: Clears the Total Hot Cycles Done and Tota;
Cold Cylcles Done fileds.