MeasurementsMXO 4 Series
234User Manual 1335.5337.02 ─ 07
"Setup/Hold time" measures and displays the setup and hold durations. "Setup/Hold
ratio" measurements return the ratio of the setup time to the sum of hold and setup
time: T
Setup
/
(T
Setup
+
T
Hold
).
The clock edge can be defined, the polarity of the data signal does not matter.
SetupHold Setup
Hold
Clock
Data
Clock slope = Positive Clock slope = Either
Clock source ← Setup/Hold time measurement
Sets the waveform used as clock in the setup/hold measurement.
Remote command:
MEASurement<mg>:SSRC on page 615
Clock slope ← Setup/Hold time measurement
Sets the edge of the clock from which the setup and hold times are measured: positive,
negative, or either of them. If "Either" is selected, the clock edges next to the data edge
are considered regardless of the clock slope.
Remote command:
MEASurement<mg>:AMPTime:CSLope on page 616
Data source ← Setup/Hold time measurement
Sets the source for the data signal.