MeasurementsMXO 4 Series
235User Manual 1335.5337.02 ─ 07
Delay measurement
The specific settings for delay measurement allow you to measure the time between
any two slopes at any reference level. Therefore, the reference levels and the slopes
must be defined for each source individually. The measurement result is negative if the
edge of the second source comes before the edge of the first source.
Slope ← Delay measurement
Sets the edge of each source, between which the delay is measured: positive, nega-
tive, or either of them.
Remote command:
MEASurement<mg>:AMPTime:DELay<n>:SLOPe on page 618
Direction ← Delay measurement
Selects the direction for counting slopes for each source: from the beginning of the
waveform, or from the end.
Remote command:
MEASurement<mg>:AMPTime:DELay<n>:DIRection on page 617
Delay to trigger measurement settings
Delay to trigger measures the time between the trigger point and the following slope of
a waveform. The delay between the trigger and the slope can be high compared to the
accuracy of the acquisition, and the trigger point can even be outside of the current
acquisition.
Automatic measurements