ApplicationsMXO 4 Series
284User Manual 1335.5337.02 ─ 07
Result table Bar chart match Description
Magnitude (dB) Max Shown by a darkened green
bar
The maximum measured value
Limit Shown by a red bar The maxim allowed value according to the
selected standard
Margin
Sets the upper limit for the display of the Y
scale.
This value can be reset with "Clear Results"
Status Pass: green bar
Dark orange bar: current
value
Light orange bar: maximum
value
Displays, if the harmonics is within the
defined limit (pass) or not (fail).
Blue outline of the row Blue bar The selected bar matches with the outlined
line in the result box.
Reference level
The power harmonic measurement requires reference levels to obtain the measure-
ment points. Power analysis uses its own reference level set, which is independent
from other measurements and is used for all power measurements. For a description
of the reference level settings, see Chapter 11.2.3, "Reference level", on page 285.
Scale
In this tab you can configure the scale for the bargraph display.
Min X Frequency
Sets the start frequency of a bargraph display. At least three bars are displayed.
Remote command:
POWer<m>:HARMonics:DISPlay:FREQuency:STARt on page 688
Max X Frequency
Sets the stop frequency of the bargraph display. The maximum value is defined by
standard and fundamental frequency.
Remote command:
POWer<m>:HARMonics:DISPlay:FREQuency:STOP on page 688
Power analysis (option R&S
MXO4-K31)