Network operation and remote controlMXO 4 Series
463User Manual 1335.5337.02 ─ 07
0
1
2
3
4
5
6
7
STBSRE
-&-
-&-
-&-
-&-
-&-
ESB
MAV
RQS/MSS
SRQ
Output
buffer
Error/event
queue
-&-
-&-
-&-
-&-
-&-
-&-
-&-
-&-
0
1
2
3
4
5
6
7
ESRESE
0
1
2
3
4
5
6
7
& = logical AND
STATus:OPERation
STATus:QUEStionable
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
CALibrating
AUToset
MEASuring
WTRIgger
TRIggered
= logical OR
PPSupply
POVerload
COVerload
TEMPerature
ADCState
GOVerloadN
NOALigndata
LIMit
MARGin
IMPrecise
PLLunlock
Figure 16-1: Overview of the status registers hierarchy
STB, SRE
The STatus Byte (STB) register and its associated mask register Service Request
Enable (SRE) form the highest level of the status reporting system. The STB pro-
vides a rough overview of the instrument status, collecting the information of the
lower-level registers.
ESR and SCPI registers
The STB receives its information from the following registers:
The Event Status Register (ESR) with the associated mask register standard
Event Status Enable(ESE).
The STATus:OPERation and STATus:QUEStionable registers which are
defined by SCPI and contain detailed information on the instrument.
Output buffer
Remote control - status reporting system

Table of Contents