Remote control commandsMXO 4 Series
512User Manual 1335.5337.02 ─ 07
Usage: Asynchronous command
Manual operation: See "Invert channel" on page 122
DISPlay:SIGNal:LABel <Signal>,<Label>
Defines and assigns a label to the specified channel waveform.
Parameters:
<Label> String with the waveform label
Setting parameters:
<Signal> C1 | C2 | C3 | C4
Usage: Asynchronous command
Manual operation: See "Label" on page 122
CHANnel<ch>:SKEW:TIME <Offset>
Sets a skew value to compensate for the delay of the measurement setup or from the
circuit specifics that the instrument cannot compensate automatically. It affects only the
selected input channel.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<Offset> Range: -100E-9 to 100E-9
Increment: 1E-12
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "Skew" on page 122
CHANnel<ch>:BANDwidth <BandwidthLimit>
Sets the bandwidth limit. The specified bandwidth indicates the range of frequencies
that the instrument can acquire and display accurately with less than 3 dB attenuation.
Frequencies above the limit are removed from the signal, and noise is reduced.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<BandwidthLimit> FULL | B2G | B1G5 | B1G | B700 | B500 | B350 | B200 | B100 |
B50 | B20
FULL
Sets the bandwidth to the maximum bandwidth of the instru-
ment. Bandwidth extension options are considered.
Acquisition and setup