Remote control commandsMXO 4 Series
511User Manual 1335.5337.02 ─ 07
Parameters:
<Position> Positive values move up the waveform, negative values move it
down.
Range: -5 to 5
Increment: 0.02
*RST: 0
Default unit: div
Usage: Asynchronous command
Manual operation: See "[Position]" on page 43
CHANnel<ch>:COUPling <Value>
Sets the connection of the channel signal, i.e. the input impedance (termination) and a
filter (coupling). The command determines what part of the signal is used for waveform
analysis and triggering.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<Value> DC | DCLimit | AC
DC
Connection with 50 Ω termination, passes both DC and AC com-
ponents of the signal.
DCLimit
Connection with 1 MΩ termination, passes both DC and AC
components of the signal.
AC
Connection with 1 MΩ termination through DC capacitor,
removes DC and very low-frequency components. The wave-
form is centered on zero volts.
Usage:
Asynchronous command
Manual operation: See "Coupling" on page 121
CHANnel<ch>:INVert <InvertChannel>
Turns the inversion of the signal amplitude on or off. To invert means to reflect the volt-
age values of all signal components against the ground level.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<InvertChannel> ON | OFF
ON: inverted waveform
OFF: normal waveform
*RST: OFF
Acquisition and setup