Remote control commandsMXO 4 Series
564User Manual 1335.5337.02 ─ 07
Usage: Asynchronous command
Manual operation: See "Source: channel states" on page 167
TRIGger:EVENt<ev>:PATTern:QUALify:DIGital:CHAN<n>:HLX <HLX>
Sets the required state for each digital channel that is used for triggering.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
<n> 0 to 15
Parameters:
<HLX> HIGH | LOW | DONTcare
*RST: DONTcare
Usage: Asynchronous command
Manual operation: See "Source: channel states" on page 168
TRIGger:EVENt<ev>:PATTern:QUALify:LOGic <StateOperator>
Defines the logic combination of the channels and their states.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<StateOperator> AND | OR
*RST: AND
Usage: Asynchronous command
Manual operation: See "State Operator" on page 167
TRIGger:EVENt<ev>:PATTern:QUALify:DIGital:LOGic <Logic>
Defines the logic combination of the channels and their states.
Suffix:
<ev>
.
1..3
Parameters:
<Logic> AND | OR
*RST: AND
Usage: Asynchronous command
Manual operation: See "State Operator" on page 167
TRIGger:EVENt<ev>:PATTern:QUALify:SOURces <StateSources>
Selects if the source of the trigger for the pattern type is an analog or digital channel.