Remote control commandsMXO 4 Series
607User Manual 1335.5337.02 ─ 07
Parameters:
<name> String with path and filename with extension *xml, *ref, *csv.
For local storage, the path is
always /home/storage/userData.
Example:
EXP:WAV:NAME
'/home/storage/userData/Export_C1.csv'
EXP:WAV:SCOP DISP
EXP:WAV:SAVE
Saves the visible waveform data of channel 1 in xml format
to /home/storage/userData/Export_C1.csv.
Usage: Asynchronous command
Manual operation: See "Save settings" on page 299
EXPort:WAVeform:SAVE
Saves the waveform to the file specified with EXPort:WAVeform:NAME.
Usage:
Event
Asynchronous command
Manual operation: See "Save settings" on page 299
EXPort:WAVeform:ABORt
Aborts a running waveform export, which was started with EXPort:WAVeform:SAVE.
Usage:
Event
Asynchronous command
EXPort:WAVeform:SCOPe <Scope>
Defines the part of the waveform record that has to be stored.
Parameters:
<Scope> DISPlay | ALL | CURSor | GATE | MANual
DISPlay
Waveform data that is displayed in the diagram.
ALL
Complete waveform, which is usually longer than the displayed
waveform.
CURSor
Data between the cursor lines if a cursor measurement is
defined for the source waveform.
GATE
Data included in the measurement gate if a gated measurement
is defined for the source waveform.
Data management