Remote control commandsMXO 4 Series
629User Manual 1335.5337.02 ─ 07
Parameters:
<UpperLevel> Range: -1E+26 to 1E+26
Increment: 0.001
*RST: 0
Default unit: V
Usage: Asynchronous command
Manual operation: See "Upper level, Middle level, Lower level" on page 225
REFLevel<rl>:RELative:HYSTeresis <HystRel>
Defines a hysteresis for the middle reference level. A rise or fall from the middle refer-
ence value that does not exceed the hysteresis is rejected as noise.
Suffix:
<rl>
.
1...4, index of the reference level set
Parameters:
<HystRel> Range: 0 to 50
Increment: 1
*RST: 10
Default unit: %
Usage: Asynchronous command
Manual operation: See "Hysteresis" on page 225
REFLevel<rl>:RELative:LOWer <LowRefLevRel>
Sets the lower relative reference level if REFLevel<rl>:RELative:MODE is set to
USER.
Suffix:
<rl>
.
1...4, index of the reference level set
Parameters:
<LowRefLevRel> Range: -100 to 200
Increment: 1
*RST: 10
Default unit: %
Usage: Asynchronous command
Manual operation: See "Upper level, Middle level,Lower level" on page 225
REFLevel<rl>:RELative:MIDDle <MidRefLevRel>
Sets the middle relative reference level if REFLevel<rl>:RELative:MODE is set to
USER.
Suffix:
<rl>
.
1...4, index of the reference level set
Automatic measurements