701User Manual 1335.5337.02 ─ 07
METer:BANDwidth <Bandwidth>
Sets the filter bandwidth.
Parameters:
<Bandwidth> B20M | B10M | B5M | B2M | B1M | B500 | B200
B20M: 20 MHz
B10M: 10 MHz
B5M: 5 MHz
B2M: 2 MHz
B1M: 1 MHz
B500: 500 kHz
B200: 200 kHz
*RST: B20M
Usage: Asynchronous command
Manual operation: See "Filter bandwidth" on page 290
METer:DVMeter<m>:ACRMs:ENABle <ACRMSResults_St>
Enables the AC RMS voltmeter measurement for the respective channel.
Suffix:
<m>
.
1 to 4, index of the analog channel
Parameters:
<ACRMSResults_St> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Measurements" on page 291
METer:DVMeter<m>:ACRMs:RESult?
Returns the result of the AC RMS voltmeter measurement.
Suffix:
<m>
.
1 to 4, index of the analog channel
Return values:
<Results_ACRMS> Range: -1E+26 to 1E+26
Increment: 0.1
*RST: 0
Default unit: V
Usage: Query only
Asynchronous command
Manual operation: See "Measurements" on page 291