TRIGger:UART:DATA <DataPattern>
Specifies the data pattern to be found on the specified trigger source. Enter the words
in MSB first bit order.
Parameters:
<DataPattern>
Usage: Asynchronous command
Manual operation: See "Value" on page 363
TRIGger:UART:DPOSition <DataPosition>
Sets the number of words before the first word of interest. These offset words are
ignored.
Parameters:
<DataPosition> Range: 1 to 4096
Increment: 1
*RST: 1
Usage: Asynchronous command
Manual operation: See "Position" on page 363
TRIGger:UART:FCONdition <DataOperator>
Selects the operator for the "Data" pattern.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
*RST: EQUal
Usage: Asynchronous command
Manual operation: See "Data" on page 363
TRIGger:UART:OPERator <DataOperator>
Sets the operator for the data pattern in the selected field of the selected frame.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
*RST: EQUal
Usage: Asynchronous command
TRIGger:UART:SOURce <Source>
Selects the transmitter or receiver line as trigger source.