Remote control commandsMXO 4 Series
786User Manual 1335.5337.02 ─ 07
SBUS<sb>:CAN:FRAMe<fr>:DLCValue?
Returns the data length code of the selected frame - the number of data bytes in the
frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<FrameDLCValue> Range: 0 to 2047
Increment: 1
*RST: 0
Usage: Query only
Asynchronous command
SBUS<sb>:CAN:FRAMe<fr>:FERCause?
Returns information on a form error, if the frame status query (SBUS<sb>:CAN:
FRAMe<fr>:STATus?) returned a form error.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<FormErrorCause> NONE | CRCDerror | ACKDerror | FSBE | RESerror
CRCDerror = CRC delimiter error
ACKDerror = ACK delimiter error
FSBE = fixed stuff bit error (CAN FD ISO only)
RESerror = reserved bit error
*RST: NONE
Usage: Query only
Asynchronous command
SBUS<sb>:CAN:FRAMe<fr>:IDSTate?
Return the states of the identifier state.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Return values:
<IdtfSt> OK | ERRor | UNDF
UNDF: Undefined
*RST: OK
Protocols