Remote control commandsMXO 4 Series
821User Manual 1335.5337.02 ─ 07
Parameters:
<Index_Min> Range: 1 to 65535
Increment: 1
*RST: 1
Usage: Asynchronous command
Manual operation: See "Edit" on page 410
SBUS<sb>:SPMI:FILTer:IMAX <Frame>,<Field>,<Data>
SBUS<sb>:SPMI:FILTer:FRAMe<fr>:FLD<fl>:IMAX <Index_Max>
Sets the end value of an index range if the operator is set to INRange.
You can set the operator with SBUS<sb>:SPMI:FILTer:FRAMe<fr>:FLD<fl>:
IOPerator.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
<fl> index of the field
Parameters:
<Index_Max> Range: 1 to 65535
Increment: 1
*RST: 65535
Usage: Asynchronous command
Manual operation: See "Edit" on page 410
SBUS<sb>:SPMI:FILTer:FRENable <Frame>,<Enabler>
SBUS<sb>:SPMI:FILTer:FRAMe<fr>:ENABle <Enable>
Enables the filtering on SPMI frames. Only the frames that match the selected filter
conditions are displayed.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
Parameters:
<Enable> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Frame type" on page 410
SBUS<sb>:SPMI:FILTer:FIENable <Frame>,<Field>,<Enabler>
SBUS<sb>:SPMI:FILTer:FRAMe<fr>:FLD<fl>:ENABle <CondEnabler>
Enables or disables the checking condition for the selected field of the selected frame.