Remote control commandsMXO 4 Series
872User Manual 1335.5337.02 ─ 07
Manual operation: See "Run single" on page 444
WGENerator<wg>:ARBGen:SAMPles?
Sets the sample rate for the arbitrary waveform.
Suffix:
<wg>
.
1...2, index of the waveform generator
Return values:
<NumSamples> Range: 0 to 128000000
Increment: 10
*RST: 0
Default unit: pts
Usage: Query only
Asynchronous command
Manual operation: See "Number of samples" on page 445
WGENerator<wg>:ARBGen:SRATe <SampleRate>
Sets the sample rate for the arbitrary waveform.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<SampleRate> Range: 1 to 312500000
Increment: 10
*RST: 1000000
Default unit: Sa/s
Usage: Asynchronous command
Manual operation: See "Sample rate" on page 445
WGENerator<wg>:ARBGen[:SOURce] <WaveformSource>
Selects the source of the arbitrary waveform.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<WaveformSource> ARBitrary | SCOPe
*RST: ARBitrary
Usage: Asynchronous command
Manual operation: See "Arb wfm source" on page 445
Waveform generator (option R&S MXO4-B6)