EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Page 872

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commandsMXO 4 Series
872User Manual 1335.5337.02 ─ 07
Manual operation: See "Run single" on page 444
WGENerator<wg>:ARBGen:SAMPles?
Sets the sample rate for the arbitrary waveform.
Suffix:
<wg>
.
1...2, index of the waveform generator
Return values:
<NumSamples> Range: 0 to 128000000
Increment: 10
*RST: 0
Default unit: pts
Usage: Query only
Asynchronous command
Manual operation: See "Number of samples" on page 445
WGENerator<wg>:ARBGen:SRATe <SampleRate>
Sets the sample rate for the arbitrary waveform.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<SampleRate> Range: 1 to 312500000
Increment: 10
*RST: 1000000
Default unit: Sa/s
Usage: Asynchronous command
Manual operation: See "Sample rate" on page 445
WGENerator<wg>:ARBGen[:SOURce] <WaveformSource>
Selects the source of the arbitrary waveform.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<WaveformSource> ARBitrary | SCOPe
*RST: ARBitrary
Usage: Asynchronous command
Manual operation: See "Arb wfm source" on page 445
Waveform generator (option R&S MXO4-B6)

Table of Contents