Remote control commandsMXO 4 Series
873User Manual 1335.5337.02 ─ 07
17.18.1.4 Sweep settings
WGENerator<wg>:SWEep[:STATe].................................................................................. 873
WGENerator<wg>:SWEep:TYPE.....................................................................................873
WGENerator<wg>:SWEep:FSTart................................................................................... 873
WGENerator<wg>:SWEep:TIME..................................................................................... 874
WGENerator<wg>:SWEep:FEND.................................................................................... 874
WGENerator<wg>:SWEep[:STATe] <Sweep>
Enables or disables the sweeping.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<Sweep> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Sweep state" on page 441
WGENerator<wg>:SWEep:TYPE <Type>
Sets the type of the sweep, a linear or logarithmic change of the frequency.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<Type> LIN | LOG
*RST: LIN
Usage: Asynchronous command
Manual operation: See "Sweep type" on page 441
WGENerator<wg>:SWEep:FSTart <StartFrequency>
Sets the start frequency of the sweep signal.
Suffix:
<wg>
.
1...2, index of the waveform generator
Parameters:
<StartFrequency> Range: 0.001 to 100000000
Increment: 0.001
*RST: 1000
Default unit: Hz
Usage: Asynchronous command
Manual operation: See "Start frequency" on page 442