Network operation and remote control
R&S
®
RTE
1049User Manual 1326.1032.02 ─ 20
Bit No. Meaning
5
ADCState
The bit is set if the signal is clipped on the upper or lower edge of the screen -overflow of the
ADC occurs (see "STATus:QUEStionable:ADCState register" on page 1050).
6 Not used
7 Not used
8
NOALigndata
This bit is set if no alignment data is available - the instrument is uncalibrated.
9 to 10 not used
11 LAMP (Low AMPlitude)
This bit is set if the magnitude of the signal is too low to get reliable measurement results. See
"STATus:QUEStionable:LAMP register" on page 1050.
12
MASK
This bit is set if a mask value is violated (see "STATus:QUEStionable:MASK register"
on page 1051
13 Not used
14 ZVCO: This bit is set if a channel overload occurs at one or more input channels of the R&S RT-
ZVC multi-channel power probe. See "STATus:QUEStionable:ZVCO register" on page 1051.
15 This bit is always 0.
STATus:QUEStionable:COVerload register
This register contains all information about overload of the channels. The bit is set if
the assigned channel is overloaded, or if an overload risk occurred (overload warning).
Table 17-4: Bits in the STATus:QUEStionable:COVerload register
Bit No. Meaning
0 Overload on CHANnel1
1 Overload on CHANnel2
2 Overload on CHANnel3
3 Overload on CHANnel4
4 Overload warning for CHANnel1
5 Overload warning for CHANnel2
6 Overload warning for CHANnel3
7 Overload warning for CHANnel4
8 Overload on external trigger input
9 Overload warning for external trigger input
STATus:QUEStionable:TEMPerature register
This register contains information about the instrument's temperature.
Remote control