Network operation and remote control
R&S
®
RTE
1050User Manual 1326.1032.02 ─ 20
Table 17-5: Bits in the STATus:QUEStionable:TEMPerature register
Bit No. Meaning
0
TEMP WARN
This bit is set if a temperature warning on channel 1, 2, 3 or 4 occurred.
1
TEMP ERRor
This bit is set if a temperature error on channel 1, 2, 3 or 4 occurred.
STATus:QUEStionable:ADCState register
This register contains all information about overflow of the ADC.
The bit is set if the assigned channel signal is clipped on the upper or lower edge of the
screen. In this case, the signal does not fit in the range of the ADC and overflow
occurs.
Table 17-6: Bits in the STATus:QUEStionable:ADCState register
Bit No. Meaning
0 CHANnel1, clipping on the upper limit
1 CHANnel1, clipping on the lower limit
2 CHANnel2, clipping on the upper limit
3 CHANnel2, clipping on the lower limit
4 CHANnel3, clipping on the upper limit
5 CHANnel3, clipping on the lower limit
6 CHANnel4, clipping on the upper limit
7 CHANnel4, clipping on the lower limit
STATus:QUEStionable:LAMP register
The LAMP (Low AMPlitude) bit is set if the magnitude of the signal is too low to get
reliable measurement results.
Table 17-7: Bits in the STATus:QUEStionable:LAMP register
Bit No. Meaning
0
MEAS1
1
MEAS2
2
MEAS3
3
MEAS4
4
MEAS5
5
MEAS6
6
MEAS7
7
MEAS8
Remote control