Remote control commands
R&S
®
RTE
1191User Manual 1326.1032.02 ─ 20
Parameters:
<ClockSource> CHAN1 | CHANNEL1 | CHAN2 | CHANNEL2 | CHAN3 |
CHANNEL3 | CHAN4 | CHANNEL4
CHAN1 = CHANnel1, CHAN2 = CHANnel2, CHAN3 = CHAN-
nel3, CHAN4 = CHANnel4
Input channel of the clock signal
*RST:  CHAN1
Usage:  Asynchronous command
TRIGger<m>:SPATtern:CSOurce:EDGE <ClockEdge>
Together with the clock level (see TRIGger<m>:SPATtern:CSOurce:LEVel
on page 1191), the clock edge defines the point in time when the state of the data sig-
nal is checked.
Suffix: 
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<ClockEdge> POSitive | NEGative | EITHer
See Chapter 18.4.3, "Slope parameter", on page 1075.
*RST:  POSitive
Usage:  Asynchronous command
TRIGger<m>:SPATtern:CSOurce:LEVel <ClockLevel>
Defines the voltage level for the clock signal.
Suffix: 
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<ClockLevel> Range:  -10  to  10
Increment:  1E-3
*RST:  0
Default unit: V
Usage:  Asynchronous command
TRIGger<m>:SPATtern:PATTern <Pattern>
The pattern contains the bits of the serial data to be found in the data stream. The
maximum length of the pattern is 128 bit.
Suffix: 
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Trigger