Getting started
R&S
®
RTE
121User Manual 1326.1032.02 ─ 20
Zoom
The zoom icon on the toolbar shows the last selected zoom type. A short tap on the
icon activates the selected zoom.
If you touch the icon and drag your finger down, a menu opens where you can select
another zoom type.
Standard zoom ← Zoom
Displays a magnified section of the diagram in an additional zoom diagram. It is a dis-
play zoom, instrument settings are not changed.
Touch and hold the zoom area to open the "Zoom" dialog box.
Remote command:
LAYout:ZOOM:ADD on page 1212
Hardware zoom ← Zoom
Changes the instrument settings - horizontal and vertical scales as well as trigger level
and offset - to display a part of the diagram in greater detail.
Coupled zoom ← Zoom
Creates a coupled zoom area and its related zoom diagram. If you change the size of
one zoom area, the size of all coupled zoom areas is changed as well.
Remote command:
LAYout:ZOOM:ADDCoupled on page 1212
Fingertip zoom ← Zoom
Magnifies the waveforms around your fingertip.
Tap the icon and put your finger on the waveform. The touched part of the waveform is
displayed in a magnifier. Drag your finger on the screen to move the magnifier. You can
change the zoom factor using the [Navigation] knob.
Search
Performs a search according to the settings in the "Search Setup" dialog box. Tap the
icon and then tap the diagram with the waveform to be searched, or drag a rectangle to
define a search gate. The search is performed on the selected waveform.
Cursor
The cursor icon shows the last selected cursor type. A short tap on the icon activates
the selected cursor.
If you touch the icon and drag your finger down, a menu opens where you can select
another cursor type: horizontal cursors, vertical cursors, or both.
Tap the diagram where you want to set the cursors, or draw a rectangle in the diagram
to position the cursor lines. The resulting cursor lines measure the selected waveform.
The results appear in the "Cursor Results" box. You can adjust the cursor source, type
and position in the result box. Move the cursor lines by dragging them in the diagram,
or by turning the navigation knob. Pressing the knob switches the parameter to be
changed.