Saves the current instrument settings in a saveset. The filename is created according
to the autonaming pattern. You can reload the saveset using the "Load saveset
(Graphical recall)" toolbar icon, or using [FILE] > "Save/Recall" > "Settings".
Save Screenshot
Saves a screenshot of the current display using the settings defined in "File" menu >
"Print Setup".
Create Report
Creates a report of the current measurement settings and results using the settings
defined in "File" menu > "Report Setup".
Clear all
Deletes all measurement results including long term measurement and statistics, all
waveforms, and the history.
Remote command:
DISPlay:CLR on page 1105
Autoset and Preset
Performs an autoset, or a preset to a default state. The icons have the same function-
ality as the corresponding keys on the front panel. They are useful when you control
the instrument remotely.
Run / Stop and Run Single
Starts and stops the continuous acquisition, or starts a defined number of acquisition
cycles. The icons have the same functionality as the corresponding keys on the front
panel. They are useful when you control the instrument remotely.
Find Trigger Level
Analyses the signal and sets the trigger level to the middle of the signal peaks.
Force Trigger
Starts an immediate single acquisition. If the acquisition is running in normal mode and
no valid trigger occurs, use "Force Trigger" to confirm that a signal is available. Then
you can use the displayed waveform to determine how to trigger on it.
Label
Defines a waveform label that names or explains the waveform. Tap the icon and then
tap the waveform to be labeled. If you tap the display background, the label is assigned
to the selected waveform. Enter the label text using the onscreen keyboard. The text is
shown in the same color as the waveform. You can drag the label to another position.
Update Ref Waveform
Copies the selected source waveform with all its settings to the reference waveform. If
the acquisition is running, the reference waveform is a snapshot. You can configure up
to four reference waveforms.
Select the required reference waveform (R1 to R4) in the toolbar menu of the icon.