Remote control commands
R&S
®
RTE
1365User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:WIDTh:DELTa <SearchName>,<WidthDelta>
SEARch:TRIGger:WIDTh:DELTa? <SearchName>
Defines a range around the given width value (see also SEARch:TRIGger:WIDTh:
WIDTh on page 1366).
Parameters:
<WidthDelta> Range: 0 to 432
Increment: 500E-12
*RST: 0
Default unit: s
Parameters for setting and query:
<SearchName> Search definition
Example: See Chapter 18.3.4.1, "Searching for a pulse of specified width",
on page 1060
Usage: Asynchronous command
SEARch:TRIGger:WIDTh:POLarity <SearchName>,<Polarity>
SEARch:TRIGger:WIDTh:POLarity? <SearchName>
Indicates the polarity of a pulse, that is the direction of the first pulse slope.
Parameters:
<Polarity> POSitive | NEGative | EITHer
See Chapter 18.4.4, "Polarity parameter", on page 1075.
*RST: POSitive
Parameters for setting and query:
<SearchName> Search definition
Usage: Asynchronous command
SEARch:TRIGger:WIDTh:RANGe <SearchName>,<RangeMode>
SEARch:TRIGger:WIDTh:RANGe? <SearchName>
Selects how the range of a pulse width is defined in relation to the width and delta (see
SEARch:TRIGger:WIDTh:WIDTh on page 1366 and SEARch:TRIGger:WIDTh:
DELTa on page 1365).
Parameters:
<RangeMode> WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers on pulses inside a given range. The range of the pulse
width is defined by "Width" and "±Delta".
OUTSide
Triggers on pulses outside a given range. The range definition is
the same as for "Within" range.
Search

Table of Contents