Remote control commands
R&S
®
RTE
1431User Manual 1326.1032.02 ─ 20
BIT7_RW
Enter 7 address bits and the R/W bit. Only available if BUS<m>:
I2C:RWBitINADdress is set.
BIT10
10-bit address
ANY
Only available for trigger type "Address + data" (TRIGger<m>:
I2C:MODE ADAT). Used to trigger on data only, regardless of
the address.
*RST: BIT7
Usage: Asynchronous command
TRIGger<m>:I2C:ACONdition <AddressOperator>
Sets the operator to set a specific address or an address range. The address values
are set with TRIGger<m>:I2C:ADDRess and TRIGger<m>:I2C:ADDTo.
Parameters:
<AddressOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
*RST: EQUal
TRIGger<m>:I2C:ADDRess <Address>
Triggers on the specified slave address, or sets the the start value of an address range
depending on the condition set with TRIGger<m>:I2C:ACONdition.
Parameters:
<Address> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076.
TRIGger<m>:I2C:ADDTo <AddressTo>
Sets the the end value of an address range if the condition is set to an address range
with TRIGger<m>:I2C:ACONdition.
Parameters:
<AddressTo> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076.
Usage: Asynchronous command
TRIGger<m>:I2C:ADOR<n>:ENABle <UseAddress>
Includes the indicated ADOR address in the "address OR" trigger condition.
Suffix:
<n>
.
1..4
Index of the address in an "address OR" condition (OR slot)
Protocols