Remote control commands
R&S
®
RTE
1489User Manual 1326.1032.02 ─ 20
BUS<m>:CAN:FDATa:PSTandard <ProtStd>
Only available for CAN FD buses. Selects whether the tested signal is an ISO CAN FD
signal or not.
Suffix: 
<m>
.
1..4
Parameters:
<ProtStd> ISO | NISO
ISO
Signals are decoded according to the the ISO CAN FD protocol.
This protocol has an additional stuff count field before the CRC
sequence.
NISO
Non-ISO. Signals are decoded according to the the Bosch CAN
FD protocol.
*RST:  ISO
Firmware/software:  FW 3.35
BUS<m>:CAN:DATA:THReshold <Threshold>
Sets a user-defined threshold value.
Alternatively, you can set the threshold according to the signal technology with
BUS<m>:CAN:TECHnology.
Suffix: 
<m>
.
1..4
Selects the serial bus.
Parameters:
<Threshold> Range:  -12  to  12
Increment:  0.1
*RST:  0
Default unit: V
BUS<m>:CAN:TECHnology <Technology>
Sets the threshold voltage as defined for various signal technologies.
Suffix: 
<m>
.
1..4
Selects the serial bus.
Parameters:
<Technology> V25 | V3 | V2 | V0 | MANual
V25
2.5 Volt (CMOS 5.0 V)
V3
3.0 Volt (CAN_H HS / CAN_L LS)
Protocols