Remote control commands
R&S
®
RTE
1607User Manual 1326.1032.02 ─ 20
BUS<m>:MILStd:POLarity............................................................................................ 1608
BUS<m>:MILStd:PRESet..............................................................................................1608
BUS<m>:MILStd:THReshold:HIGH................................................................................1608
BUS<m>:MILStd:THReshold:LOW.................................................................................1609
BUS<m>:MILStd:SOURce <SourceData>
Sets the channel for the signal source.
Suffix: 
<m>
.
1..4
Selects the serial bus.
Parameters:
<SourceData> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4
*RST:  C1W1
BUS<m>:MILStd:MAXResponse:BITS 
<MaxResponseTime>
Sets the value for the maximum response time.
See also: BUS<m>:MILStd:MAXResponse:SELect.
Suffix: 
<m>
.
1..4
Parameters:
<MaxResponseTime> Range:  2E-6  to  262E-6
Increment:  100E-9
*RST:  14E-6
Default unit: s
BUS<m>:MILStd:MAXResponse:SELect <MaxRespSel>
Enables the detection of the maximum response time between the last bit of a word
and the following status word sync during decoding.
To specify the maximum response time, use BUS<m>:MILStd:MAXResponse:BITS
on page 1607.
Suffix: 
<m>
.
1..4
Parameters:
<MaxRespSelect> ON | OFF
*RST:  ON
BUS<m>:MILStd:MINGap:SELect <MinGapSelect>
Enables the detection of the minimum idle time between the last bit of a message and
the following command word sync during decoding.
Protocols