Remote control commands
R&S
®
RTE
1612User Manual 1326.1032.02 ─ 20
Parameters:
<RTAPatternMax> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MILStd:CMD:CCONdition <WordCntOperator>
Sets the operator to set a specific data word count or mode code pattern.
Parameters:
<WordCntOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one pattern to
be set with TRIGger<m>:MILStd:CMD:CMIN.
INRange | OORange
In range/Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:MILStd:CMD:CMIN and
TRIGger<m>:MILStd:CMD:CMAX.
*RST: EQUal
TRIGger<m>:MILStd:CMD:CMAX <WordCntPattMax>
Sets the end value of a data word count/mode code pattern if TRIGger<m>:MILStd:
CMD:CCONdition is set to INRange or OORange.
Parameters:
<WordCntPattMax> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MILStd:CMD:CMIN <WordCntPattMin>
Specifies a data word count/mode code pattern or sets the the start value of a pattern
range.
Parameters:
<WordCntPattMin> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MILStd:CMD:SCONdition <SubaddrOperator>
Sets the operator to set a specific subaddress/mode pattern.
Parameters:
<SubaddrOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
Protocols