Remote control commands
R&S
®
RTE
1695User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:HBTO:DATA:DMAX <SearchName>,<DataMax>
SEARch:TRIGger:HBTO:DATA:DMAX? <SearchName>
Sets the end value of the data range if SEARch:TRIGger:HBTO:DATA:DCONdition
is set to INRange or OORange.
Parameters:
<DataMax> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:HBTO:DATA:ICONdition <SearchName>,<DataIdxOperator>
SEARch:TRIGger:HBTO:DATA:ICONdition? <SearchName>
Sets the condition for the index. You can define an exact value or a value range.
Parameters:
<DataIdxOperator> EQUal | LTHan | LETHan | GTHan | GETHan | INRange |
RANGe
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding SEARch:TRIGger:HBTO:
DATA:IMIN command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with SEARch:TRIGger:HBTO:DATA:IMIN and
SEARch:TRIGger:HBTO:DATA:IMAX.
*RST:
INRange
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:HBTO:DATA:IMIN <SearchName>,<DataIndexMin>
SEARch:TRIGger:HBTO:DATA:IMIN? <SearchName>
Specifies the index bit pattern, or sets the start value of a pattern range.
Parameters:
<DataIndexMin> Range: 1 to 0
Increment: 1
*RST: 1
Parameters for setting and query:
<SearchName>
Protocols