Remote control commands
R&S
®
RTE
1801User Manual 1326.1032.02 ─ 20
ACK
Sets the trigger to the ACK field (acknowledgment of error-free
data packet).
NAK
Sets the trigger to the NAK field (non-acknowledgment, no suc-
cessful data transmission).
STALl
Sets the trigger to the STALL field (endpoint is halted or a con-
trol pipe request is not supported).
NYET
Sets the trigger to the NYET field (no response yet from
receiver).
*RST:
ACK
TRIGger<m>:USB:PATT
<PayloadMuster>
Specifies the payload data pattern that is to be searched.
Parameters:
<PayloadMuster>
TRIGger<m>:USB:PCONdition <OperatorPort>
Sets the operator to set a specific port number or a port number range. The port num-
ber values are set with TRIGger<m>:USB:PMIN and TRIGger<m>:USB:PMAX.
Parameters:
<OperatorPort> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one port number
to be set using TRIGger<m>:USB:PMIN.
INRange | OORange
In range, out of range. These conditions require a port number
range to be set using TRIGger<m>:USB:PMIN and
TRIGger<m>:USB:PMAX.
*RST: OFF
TRIGger<m>:USB:PMIN <Port>
Specifies the port number, or sets the the start value of a port number range.
Parameters:
<Port>