Remote control commands
R&S
®
RTE
1835User Manual 1326.1032.02 ─ 20
Frame
number
Frame name Field
number
Field name
10 Req Chunk
11 Data Size
12 DATA
TRIGger<m>:USBPd:ERENable....................................................................................1835
TRIGger<m>:USBPd:ERRor<n>:ENABle....................................................................... 1835
TRIGger<m>:USBPd:FRENable....................................................................................1836
TRIGger<m>:USBPd:FRAMe<n>:ENABle...................................................................... 1836
TRIGger<m>:USBPd:FIENable..................................................................................... 1836
TRIGger<m>:USBPd:FRAMe<n>:FLD<o>:ENABle..........................................................1836
TRIGger<m>:USBPd:BIT..............................................................................................1837
TRIGger<m>:USBPd:FRAMe<n>:FLD<o>:BIT................................................................1837
TRIGger<m>:USBPd:DMAX..........................................................................................1837
TRIGger<m>:USBPd:FRAMe<n>:FLD<o>:DMAX........................................................... 1837
TRIGger<m>:USBPd:DMIN...........................................................................................1837
TRIGger<m>:USBPd:FRAMe<n>:FLD<o>:DMIN.............................................................1837
TRIGger<m>:USBPd:DOPerator................................................................................... 1838
TRIGger<m>:USBPd:FRAMe<n>:FLD<o>:DOPerator..................................................... 1838
TRIGger<m>:USBPd:IMAX...........................................................................................1838
TRIGger<m>:USBPd:FRAMe<n>:FLD<o>:IMAX.............................................................1838
TRIGger<m>:USBPd:IMIN............................................................................................ 1839
TRIGger<m>:USBPd:FRAMe<n>:FLD<o>:IMIN..............................................................1839
TRIGger<m>:USBPd:IOPerator.....................................................................................1839
TRIGger<m>:USBPd:FRAMe<n>:FLD<o>:IOPerator.......................................................1839
TRIGger<m>:USBPd:FRAMe<n>:FLD<o>:BIT................................................................1839
TRIGger<m>:USBPd:ERENable <ErrorName>, <Enabler>
TRIGger<m>:USBPd:ERRor<n>:ENABle <Enable>
Enables or disables the checking condition for a specific error in the selected field of
the selected frame.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
<n> *
Specifies the error number.
Parameters:
<Enable> ON | OFF
*RST: ON
Example:
TRIGger:USBPd:ERRor1:ENABle ON
Enables CRC Error.
TRIGger:USBPd:ERENable "Length Error",ON
Enables Length Error.