Remote control commands
R&S
®
RTE
1902User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
0..15
Number of the digital channel.
Channel groups: 0..3; 4..7; 8..11; 12..15
Parameters:
<Value> Range: -8.0 to 8.0
Increment: 200.0e-12
*RST: 0
Default unit: V
DIGital<m>:THCoupling
<State>
Sets the threshold and the hysteresis for all digital channels of parallel bus 1 to the
same value.
The command BUS<m>:PARallel:THCoupling is used to set all buses.
Suffix:
<m>
.
0..15
The suffix is irrelevant.
Parameters:
<State> ON | OFF
DIGital<m>:HYSTeresis <Hysteresis>
Defines the size of the hysteresis to avoid the change of signal states due to noise for
the channel group to which the indicated digital channel belongs.
The setting affects only the settings of the first MSO bus (Bus1). You can set the hyste-
resis for all buses with BUS<m>:PARallel:HYSTeresis<n>.
See also: DIGital<m>:THCoupling
Suffix:
<m>
.
0..15
Number of the digital channel
Channel groups: 0..3; 4..7; 8..11; 12..15
Parameters:
<Hysteresis> MAXIMUM | MAXimum | ROBUST | ROBust | NORMAL |
NORMal
MAXIMUM = MAXimum
Maximum value that is possible and useful for the signal and its
settings
ROBUST = ROBust
Different hysteresis values for falling and rising edges to avoid
an undefined state of the trigger system.
NORMAL = NORMal
The instrument sets a value suitable for the signal and its set-
tings.
Mixed signal option (MSO, R&S RTE-B1)