Waveform analysis
R&S
®
RTE
294User Manual 1326.1032.02 ─ 20
Vertical
Vertical reference scaling changes the vertical display of the reference waveform inde-
pendently of the settings of the source waveform.
Enable ← Vertical
If enabled, the vertical offset is applied to the reference waveform.
Remote command: 
REFCurve<m>:RESCale:VERTical:STATe on page 1223
Offset ← Vertical
Moves the reference waveform vertically. Enter a value with the unit of the waveform.
Like vertical offset of a channel waveform, the offset of a reference waveform is sub-
tracted from the measured value. Negative values shift the waveform up, positive val-
ues shift it down.
Note: As for all waveforms, a vertical offset of a reference waveform can be set using
the vertical [POSITION / REF POINT] knob. This offset is independent from the refer-
ence scaling offset, which is described here. If both offsets are set, their values are
added up.
Remote command: 
REFCurve<m>:RESCale:VERTical:OFFSet on page 1223
Horizontal
Horizontal reference scaling changes the horizontal display of the reference waveform
independently of the settings of the source waveform and of the diagram settings.
Enable ← Horizontal
If enabled, the horizontal offset is applied to the reference waveform.
Remote command: 
REFCurve<m>:RESCale:HORizontal:STATe on page 1223
Position ← Horizontal
Moves the waveform horizontally. Enter a value with a time unit suitable for the time
scale of the diagram. Positive values shift the waveform to the right, negative values
shift it to the left.
Remote command: 
REFCurve<m>:RESCale:HORizontal:OFFSet on page 1224
Reference waveforms