Protocol analysis
R&S
®
RTE
620User Manual 1326.1032.02 ─ 20
Bus plus 2.5 V
Threshold high
Threshold low-A
+A
Bus minus
Idle 0 1 0 1 Idle
Single-endedDifferential
BP - BM
+2A
-2A
0 V
Threshold high
Threshold low
2.5 V
Threshold high
Threshold low
For measurements inside the FlexRay node (with "Source type" = "Logic"), each line
requires its threshold level.
There are two ways to set the thresholds: selection of a predefined value, or direct
entry of a value.
"Preset thresholds"
Selects default threshold voltages from a list. The predefined values depend on the
selected source type. The value is set to "Manual" if at least one threshold was
entered directly.
"Threshold high" and "Threshold low"
Upper and lower levels for single-ended or differential source types. You can enter
the values directly in the fields.
"Threshold data" and "Threshold enable"
Levels for data and enable line if there is logic source type. You can enter the val-
ues directly in the fields.
Note: If the sources are digital channels, the same threshold values are used for the
parallel and the serial buses. You can set the thresholds either in the parallel bus con-
figuration or in the serial bus configuration.
Remote command:
BUS<m>:FLXRay:PRSingle on page 1560
BUS<m>:FLXRay:PRDiff on page 1561
BUS<m>:FLXRay:PRLogic on page 1561
BUS<m>:FLXRay:THReshold<n> on page 1559
BUS<m>:FLXRay:THData on page 1560
BUS<m>:FLXRay:THENable on page 1560
BUS<m>:SETReflevels on page 1421
Polarity
Selects the wire on which the bus signal is measured if there is "Single-ended" mea-
surement: "Bus plus" or "Bus minus". The setting affects the digitization of the signal.
FlexRay (option R&S
RTE-K4)