Protocol analysis
R&S
®
RTE
684User Manual 1326.1032.02 ─ 20
Remote command:
TRIGger<m>:ARINc:DATA:CONDition on page 1636
TRIGger<m>:ARINc:DATA:MIN on page 1636
TRIGger<m>:ARINc:DATA:MAX on page 1637
SDI / SSM ← Label + Data
Sets the values for the source/destination identifier (SDI) and the sign/status matrix
(SSM) bits.
Remote command:
TRIGger<m>:ARINc:SDI on page 1639
TRIGger<m>:ARINc:SSM on page 1639
Error Conditions
Specifies the error conditions to be triggered on.
Coding error ← Error Conditions
Triggers on a coding error.
Remote command:
TRIGger<m>:ARINc:ERRor:CODing on page 1637
Parity Error ← Error Conditions
Checks the parity and triggers if the parity is even.
Remote command:
TRIGger<m>:ARINc:ERRor:PARity on page 1637
Timing: Min gap, Max gap ← Error Conditions
Defines the idle time between two words, which is needed for word synchronization.
The beginning of the first bit after the gap marks the start of a new word.
You can define a minimum idle time "Min gap", and/or a maximum time "Max gap". The
standard defines a minimum of 4-bit times to separate two subsequent words.
Timing settings are relevant for protocol configuration and error trigger.
If "Min gap" and/or "Max gap" are enabled, the instrument detects the specified gaps
during decoding. If the trigger type "Error condition" is selected in addition, the instru-
ment triggers when the gap is shorter or longer than the specified gaps, respectively.
ARINC 429 (option R&S
RTE-K7)