Loading...
Protocol analysis
R&S
®
RTE
700User Manual 1326.1032.02 ─ 20
"10BASE-T"
Selects the Ethernet protocol variant 10BASE-T (standard data rate
10 Mbit/s).
Figure 13-54: Ethernet 10BASE-T protocol configuration (here with inverted
polarity)
"100BASE-TX"
Selects the Ethernet protocol variant 100BASE-TX, which provides
100 Mbit/s use data rate. Due to 4b/5b encoding, the raw data rate on
the line is 125 Mbit/s. This value is used by R&S RTE-K8 as the bit
rate default for 100BASE-TX.
Figure 13-55: Ethernet 100BASE-TX protocol configuration
Remote command:
BUS<m>:ETHernet:VARiant on page 1649
Data Pattern
Specifies the data pattern that is to be triggered on.
Remote command:
TRIGger<m>:ETHernet:PATTern on page 1652
Bit rate
Defines the transmission speed setting for the data signal:
10BASE-T: default bit rate 10 Mbps
100BASE-TX: default bit rate 125 Mbps
Ethernet 10BASE-T and 100BASE-TX (option R&S
RTE-K8)