Instrument Function
R&S
®
SMU200A
260Operating Manual 1007.9845.32 ─ 15
The "Dwell Time" set by the user is used as the step time of the sweep. The effective net
dwell time is shorter, reduced by the setting time. This setting time may be greater than
the time specified in the data sheet.
Note:
It is recommended to switch off the GUI update for optimum sweep performance espe-
cially with short dwell times (see chapter 5.2.3.4, "Gui Update", on page 132).
SCPI command:
[:SOURce<hw>]:SWEep:POWer:DWELl on page 781
Global Trigger/Clock Settings - Level Sweep
Open the "Global Trigger/Clock Settings" menu. This menu is used among other things
for setting the polarity of the active slope of an applied instrument trigger
In the case of two-path instruments this setting is valid for both paths. It affects the INST
TRIG input (BNC connector at the rear of the instrument, instrument trigger for Path A)
and the INST TRIG B input of the AUX I/O interface at the rear of the instrument, instru-
ment trigger for Path B).
See chapter 5.2.3.10, "Global Trigger/Clock/External Input Settings", on page 137.
User Marker / AUX I/O - Global Settings
Calls the menu for configuring the AUX I/O and user interfaces (seechapter 5.2.3.12,
"User Marker / AUX I/O Settings", on page 140 ).
The pin assignment of the AUX I/O interface is shown in the "User Marker / AUX I/O
Configuration" menu; the assignment of the USER pins 2...4 and USER 1 BNC connector
can be changed.
With two-path instruments, the current assignment to the respective path is given in
addition to the pin assignment.|attribute=2nd_path:Release
Attenuator Mode - Level Sweep with High Power Option
(High Power Option only)
Sets the ranges of level settings for the level sweep. The sweep is either performed in
the low level (electronically switching attenuator) or in the high level (relay-switched high-
power-bypass) ranges.
Parameter "Attenuator Mode" is only available when the instrument is equipped with the
high-power output option,R&S SMU-B31 or R&S SMU-B36.
Note: An error message is generated and the best possible level is set if the level setting
for the current sweep step is outside the selected range.
"Normal "
The level settings for the level sweep are made in the area of the elec-
tronically switching attenuator. The high level ranges are not available.
"High Power "
The level settings for the level sweep are made in the area of the option.
Only the high level range is available.
SCPI command:
[:SOURce<hw>]:SWEep:POWer:AMODe on page 780
RF Signal and Analog Modulations - RF Block