Instrument Function
R&S
®
SMU200A
259Operating Manual 1007.9845.32 ─ 15
If "Step" is set, the level for the next level step of the sweep is entered here.
SCPI command:
[:SOURce<hw>]:POWer:MANual on page 768
Shape - RF Level Sweep
Selects the cycle mode for a sweep sequence (shape) .
"Sawtooth"
One sweep runs from the start level to the stop level. The subsequent
sweep starts at the start level again, i.e. the shape of sweep sequence
resembles a sawtooth.
"Triangle"
One sweep runs from start to stop level and back, i.e. the shape of the
sweep resembles a triangle. Each subsequent sweep starts at the start
level again.
SCPI command:
[:SOURce<hw>]:SWEep:POWer:SHAPe on page 782
Step - Level Sweep
Sets the step width for the individual sweep steps. This entry is effective for all sweep
modes.
With the level sweep, the logarithmic step width is a constant fraction of the current level.
This fraction is added to the current level. The logarithmic step width is entered in dB.
SCPI command:
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic] on page 783
Step - Level Sweep
Sets the step width for the individual sweep steps. This entry is effective for all sweep
modes.
With the level sweep, the logarithmic step width is a constant fraction of the current level.
This fraction is added to the current level. The logarithmic step width is entered in dB.
SCPI command:
[:SOURce<hw>]:SWEep:POWer:STEP[:LOGarithmic] on page 783
Dwell Time - Level Sweep
Enters the dwell time and determines the duration of the individual sweep steps.