Instrument Function
R&S
®
SMU200A
264Operating Manual 1007.9845.32 ─ 15
"Extern Start/
Stop"
Sets an automaticly repeated sweep cycle that is started, stopped and
restartet by subsequent external trigger events.
The first external trigger signal starts the sweep (Start).
The next external trigger signal stops the sweep at the current frequency
(Stop).
The third external trigger signal starts the sweep at the start frequency
(Start).
If a different sweep mode was activated prior to the "Extern Start/Stop"
mode, the current sweep is stopped and the "Extern Start/Stop" sweep
starts at the start frequency when triggered.
The "Reset" button resets the sweep to the start frequency.
The external trigger signal is input at the rear of the instrument (BNC
connector INST TRIG).
Example:
SOUR:LFO:SWE:FREQ:MODE AUTO
TRIG0:SWE:SOUR EAUT
SOUR:LFO:FREQ:MODE SWE (External trigger)
SCPI command:
[:SOURce]:LFOutput:SWEep[:FREQuency]:MODE on page 743
:TRIGger<hw>[:SWEep]:SOURce on page 802
[:SOURce]:LFOutput:FREQuency:MODE on page 740
Execute Single Sweep - LF Sweep
Triggers the sweep manually. A manual sweep can only be triggered if Mode Single is
selected.
SCPI command:
[:SOURce]:LFOutput:SWEep[:FREQuency]:EXECute on page 742
Reset Sweep - LF Sweep
Resets the sweep. The start frequency is set and the next sweep starts from there.
SCPI command:
[:SOURce<hw>]:SWEep:RESet[:ALL] on page 784
Start Freq - LF Sweep
Sets the start frequency.
SCPI command:
[:SOURce]:LFOutput:FREQuency:STARt on page 741
Stop Freq - LF Sweep
Sets the stop frequency.
SCPI command:
[:SOURce]:LFOutput:FREQuency:STARt on page 741
Current Freq - LF Sweep
Displays the current frequency.
RF Signal and Analog Modulations - RF Block