Remote Control Commands
R&S
®
SMU200A
742Operating Manual 1007.9845.32 ─ 15
[:SOURce]:LFOutput:FREQuency:STOP <Stop>
This command sets the stop frequency for the LF sweep.
Parameters:
float
0.1 Hz to 1 MHz
0.1 Hz
50 kHz
Hz
<Stop>
Range:
Increment:
*RST:
Default unit:
LFO:FREQ:STOP 200 kHz
sets the stop frequency for the LF sweep to 200 kHz.
LFO:FREQ:STAR 100 kHz
sets the start frequency for the LF sweep to 100 kHz.
Example:
SCPI conformUsage:
[:SOURce]:LFOutput[:STATe] <State>
The command activates/deactivates the LF output.
Parameters:
0|1|OFF|ON
OFF
<State>
*RST:
LFO ON
activates the LF output. The settings under LFO:FREQ and
LFO:SWE become effective.
Example:
[:SOURce]:LFOutput:SWEep[:FREQuency]:DWELl <Dwell>
The command sets the dwell time for each frequency step of the sweep.
Parameters:
float
2 ms to 10 s
0.1 ms
10 ms
<Dwell>
Range:
Increment:
*RST:
LFO:SWE:DWEL 20 ms
sets a dwell time of 20 ms.
Example:
[:SOURce]:LFOutput:SWEep[:FREQuency]:EXECute
The command immediately starts an LF sweep. The MODE setting
([:SOURce]:LFOutput:SWEep[:FREQuency]:MODE) determines which sweep is
executed, e.g. SOURce:LFOutput:SWEep:FREQuency:MODE STEP. The command
corresponds to the manual-control Execute Single Sweep - LF Sweep.
SOURce Subsystem

Table of Contents