Instrument Function
R&S
®
SMU200A
427Operating Manual 1007.9845.32 ─ 15
The carrier power that is set with Power + n* Power Step must be within the valid value
range -80 dB to 0 dB.
SCPI command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:POWer:STEP on page 675
Phase Start
Sets the phase of the starting carrier. This setting is only available for "Optimize Crest
Factor Mode" = Off.
SCPI command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:PHASe[:STARt] on page 675
Phase Step
Sets the width of the step with which the phase will be changed from carrier to carrier.
The phase that is set with Phase + n* Phase Step must be within the valid value range
0° to 360°.
This setting is only available for "Optimize Crest Factor Mode" = Off.
SCPI command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:PHASe:STEP on page 674
Accept
Adopts the carrier range setting into the table ("Carrier Table").
SCPI command:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:EXECute on page 674
Carrier Table and Carrier Graph
The lower part of the menu is used to open submenus for configuring carrier tables and
for setting triggers, markers and clocks.
Carrier Table
Calls the table for configuring individual carriers. This configuration can be checked with
the aid of the "Carrier Graph".
The table displays the settings of all available carriers. Carriers in the On state are high-
lighted. All carrier parameters can be edited in the table.
Baseband Signal - Baseband Block