EasyManua.ls Logo

Rohde & Schwarz SMU200A - Page 448

Rohde & Schwarz SMU200A
888 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMU200A
428Operating Manual 1007.9845.32 ─ 15
The Multi Carrier CW signal is only computed when the "Accept" button is pressed.
Whenever the table contains settings that have not yet been adopted with the "Accept"
button, the background is yellow.
Note: The phase/deg settings are only valid if optimization of the crest factor is disabled
("Optimize Crest Factor" = Off).
"No."
This is the carrier index.
"State"
Switch a carrier on/off
"Power"
Sets the power of a carrier.
"Phase"
Sets the starting phase of a carrier.
"Accept"
Transfer the settings in the carrier table into the instrument.
SCPI command:
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:STATe on page 670
[:SOURce<hw>]:BB:MCCW:CARRier:STATe on page 672
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:POWer on page 669
[:SOURce<hw>]:BB:MCCW:CARRier:POWer on page 671
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:PHASe on page 669
[:SOURce<hw>]:BB:MCCW:CARRier:PHASe on page 671
Carrier Graph
Calls a graphical representation of the chosen carrier configuration.
The carriers are on the X-axis and the colored bars represent those carriers which are in
the On state. Power is on the Y-axis, and the height of the bars corresponds to the chosen
power of each individual carrier.
SCPI command:
n.a.
Trigger/Marker
Calls the "Trigger/Marker" dialog (see chapter 5.8.5.1.5, "Trigger/Marker/Clock Set-
tings", on page 429).
Baseband Signal - Baseband Block