R&S
®
SMU200A
585Operating Manual 1007.9845.32 ─ 15
[:SOURce<hw>]:BB:ARB:TRIGger:EXTernal:SYNChronize:OUTPut <Output>
(enabled for "Trigger Source" External)
Enables/disables output of the signal synchronous to the external trigger event.
Parameters:
0|1|OFF|ON
ON
The signal calculation starts simultaneously with the external
trigger event but because of the instrument’s processing time the
first samples are cut off and no signal is outputted. After elapsing
of the internal processing time, the output signal is synchronous
to the trigger event.
OFF
The signal output begins after elapsing of the processing time and
starts with sample 0, i.e. the complete signal is outputted.
This mode is recommended for triggering of short signal
sequences with signal duration comparable with the processing
time of the instrument.
ON
<Output>
*RST:
BB:ARB:TRIG:SOUR EXT
sets external triggering.
BB:ARB:TRIG:EXT:SYNC:OUTP ON
enables synchronous output to external trigger
Example:
[:SOURce<hw>]:BB:ARB:TRIGger:OBASeband:DELay <Delay>
The command specifies the trigger delay (expressed as a number of samples) for trig-
gering by the signal from the second path (two-path instruments only).
Parameters:
float
0 to 2^32-1
1
0
sample
<Delay>
Range:
Increment:
*RST:
Default unit:
BB:ARB:TRIG:SOUR OBAS
sets for path A the internal trigger executed by the signal from the
second path (path B).
BB:ARB:TRIG:DEL 200
sets a delay of 200 samples for the trigger.
Example:
SOURce Subsystem