Remote Control Commands
R&S
®
SMU200A
586Operating Manual 1007.9845.32 ─ 15
[:SOURce<hw>]:BB:ARB:TRIGger:OBASeband:INHibit <Inhibit>
The command specifies the number of samples by which a restart is to be inhibited fol-
lowing a trigger event.
This command applies only for triggering by the second path.
Parameters:
float
0 to 2^32
1
0
sample
<Inhibit>
Range:
Increment:
*RST:
Default unit:
BB:ARB:TRIG:SOUR OBAS
sets for path A the internal trigger executed by the signal from the
second path (path B).
BB:ARB:TRIG:INH 200
sets a restart inhibit for 200 samples following a trigger event.
Example:
[:SOURce<hw>]:BB:ARB:TRIGger:OUTPut:DELay:FIXed <Fixed>
The command restricts the marker delay setting range to the dynamic range. In this range
the delay can be set without restarting the marker and signal. If a delay is entered in
setting ON but is outside this range, the maximum possible delay is set and an error
message is output.
The setting always affects every marker.
Parameters:
0|1|OFF|ON
OFF
<Fixed>
*RST:
BB:ARB:TRIG:OUTP:DEL:FIX ON
restricts the marker signal delay setting range to the dynamic
range.
Example:
[:SOURce<hw>]:BB:ARB:TRIGger:OUTPut<ch>:DELay <Delay>
The command defines the delay between the signal on the marker outputs and the start
of the signals, expressed in terms of samples.
Command :BB:ARB:TRIGger:OUTPut:DELay:FIXed ON can be used to restrict the
range of values to the dynamic range, i.e. the range within which a delay of the marker
signals can be set without restarting the marker and signal.
SOURce Subsystem