External Component Replacement
8-6 Service Manual
Requirements:
• WarmTouch power cord (page A-2)
WARNING:
Disconnect the warming unit's power cord from the AC power source before
performing this procedure.
To replace the power cord:
1. If the power cord is secured around the back of the unit, unwind it completely.
2. Position the warming unit on its front so the bottom of the unit is accessible
(Figure 8-4 on page 8-5).
3. Disengage the power cord from its routing bracket and remove it from the socket.
Caution:
The institution should follow local governing ordinances and recycling
instructions regarding disposal or recycling of filter and device
components or end of life of the product.
4. Dispose of the power cord as required by WEEE Directive.
5. Insert the new power cord into the socket, making sure that it is fully seated.
Note: Do not route the power cord through the metal hook on the filter enclo-
sure.
6. Press the power cord into the routing bracket until it is completely seated. If nec-
essary, use a screwdriver to carefully spread the sides of the bracket while insert-
ing the cord.
7. Perform the Power-On Test on page 7-3.
8. Wrap the power cord around the back of the unit, and secure it with the attached
strap (Figure 8-5).