Safety Information
Service Manual 1-3
WARNING:
No modification of this equipment is allowed. Modification can result in
death, injury, or property damage.
WARNING:
Before attempting to open or disassemble the warming unit, disconnect the
power cord from the AC power source.
WARNING:
Do not operate the warming unit with the filter enclosure removed.
1.2.3 Cautions (Service)
Caution:
Do not spray, pour, or spill any liquid on the warming unit, its accessories,
connectors, switches, or openings in the case.
Caution:
The HEPA filter must
be replaced every 2,000 hours of operation or 365 days,
whichever comes first.
Caution:
Operation of the warming system may affect or be affected by other devices
in the vicinity due to electromagnetic interference (EMI). If interference
occurs, try increasing the distance between devices, repositioning the cabling,
or plugging the devices into separate outlet circuit branches. Reference
Electromagnetic Compatibility (EMC) on page 13-4 for additional guidance.
Caution:
The institution should follow local governing ordinances and recycling
instructions regarding disposal or recycling of filter and device components or
end of life of the product.
Caution:
Observe ESD (electrostatic discharge) precautions when servicing the
warming unit.