Front Enclosure Replacement
Service Manual 12-3
WARNING:
Disconnect the warming unit's power cord from the AC power source before
performing this procedure.
To replace the front enclosure:
1. Follow the procedure Separating the Front and Rear Enclosures on page 9-2.
2. Remove the equipotential stud from the enclosure as follows (Figure 12-2):
a. Remove the top nut and lock washer from the threaded end of the stud.
b. Lift the power supply ground cable terminal off the stud.
c. Remove the next nut and lock washer from the stud.
d. Lift the AC power inlet ground cable terminal off the stud.
e. Pull the stud and ground washer out of the enclosure from the bottom.
f. Save all components for installation in the new enclosure.
Figure12-2.Equipotential Stud Connections
1 Power Supply Ground Cable
(Nut, Lock Washer,Terminal)
2 AC Power Inlet Ground Cable
(Nut, Lock Washer, Terminal)