Operation
4-6 Service Manual
Figure4-3.Warming System Set to 40°C (Medium)
Note that when you change the temperature setting, an amber arrow appears
next to the current temperature while it adjusts. When the temperature is
within approximately 1.5°C of the new setting, the arrow disappears.
4.6.1 Using Boost Mode
Boost mode allows rapid warming of patients. Air is generated at 47°C
(116.6°F) for 45 minutes. After 45 minutes, the setting automatically changes
to 45°C (High).
To use Boost mode, press the 47°C (Boost Mode) key. A countdown timer
shows the time remaining before the temperature switches to the 45°C (High)
setting. Boost mode sessions can be repeated as necessary.
Figure4-4.Warming System Set to Boost Mode