Product Specifications
13-2 Service Manual
13.3 Electrical Requirements
13.4 Environmental Specifications
13.4.1 Operating
13.4.2 Shipping and Storage
Note:
Before operating the warming unit after it has been shipped or stored, allow the unit
to acclimate to the operating environment for at least 60 minutes.
Power Requirements 100 to 240 V AC
Max. current at 100 V = 8 A
Max. current at 240 V = 5 A
Input Frequency 50/60 Hz ±1 Hz
Fuse (x2) Littlefuse (mfr.) 0218010; 250 V, 10 A
Temperature 18°C to 28°C (64.4°F to 82.4°F)
Relative Humidity 15% to 85% (non-condensing)
Temperature -40ºC to 70ºC (-40ºF to 158ºF)
Relative Humidity 10% to 95% (non-condensing)
Atmospheric Pressure 12 kPa to 106 kPa (90 mmHg to 795 mmHg)