Covidien WarmTouch - Figure 9-3 Enclosure Screws

Covidien WarmTouch
314 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Accessing the Inside of the Warming Unit
9-4 Service Manual
Figure9-3.Enclosure Screws
5. Supporting both enclosures, reposition the unit on its back.
Caution:
The front and rear enclosures are connected by cables. When separating
the enclosures, do not apply stress to the cables.
6. Carefully lift the front enclosure at the nozzle hook and rotate it away from the
rear enclosure. Note that if you did not remove the power cord previously, you will
need to disengage it from its socket while rotating the front enclosure. The power
cord can remain with the rear enclosure.
7. Set the front enclosure next to the rear enclosure, taking care not to apply stress
to the connecting cables (Figure 9-1 on page 9-2).