Performance Specifications
Service Manual 13-3
13.5 Performance Specifications
13.6 Product Compliance
Maximum Contact Surface Temperature 44.1°C (111.4°F)
Average Time for Contact Surface Temperature
to Rise from 23°C ±2°C to 37°C
6 minutes
Average Time for Temperature of Air Exiting
the Hose to Rise from 23°C ±2°C to 37°C
< 1 minute
Accuracy of Displayed Temperature ±1.0°C (air entering hose)
Automatic Temperature Stepdown
(Boost to High Temperature)
After 45 minutes of continuous use,
blower will step down from the Boost
to High setting.
Thermal Protection Threshold Thermostat (internal):
49°C to 55°C (120°F to 131°F)
Average Alarm Level 56 dB @ 1 meter
Equipment Classification IEC 60601-1:2005, EN 60601-1:2006
ANSI/AAMI ES60601-1:2005
CAN/CSA C22.2 No 60601-1:08
IEC 80601-2-35:2009
EN 80601-2-35:2010
Protection Type Class I
Degree of Protection Type BF - Applied part
Mode of Operation Continuous
Electromagnetic Compatibility IEC/EN 60601-1-2:2007, Class A Emissions,
Non Life-Support Equipment. IEC/EN 80601-2-35;
Hazard Free to 10 V/m.
Protection Against Ingress of Fluids Ordinary