Operation
4-2 Service Manual
4.3 Applying Power
WARNING:
Possible electric shock hazard. Grounding reliability can be achieved only
when the warming unit is connected to a suitable mains outlet with
protective earth grounding.
To apply power to the warming unit:
1. Locate a hospital-grade or suitable mains outlet with protective earth grounding.
Ensure that the power outlet is easily accessible; disconnection from the outlet is
the only way to completely remove power from the warming unit.
2. Plug the warming unit’s power cord into the outlet. The status LED illuminates
amber to indicate that the warming unit is in Standby mode. The fan and heater
are off, and the display is blank.